['Drug and Alcohol Testing - DOT']
['Drug testing - Motor Carrier']
07/21/2022
...
As a laboratory, when you conduct validity testing under §40.89, you must conduct it in accordance with the requirements of this section.
(a) You must determine the creatinine concentration on each primary specimen. You must also determine its specific gravity if you find the creatinine concentration to be less than 20 mg/dL.
(b) You must determine the pH of each primary specimen.
(c) You must perform one or more validity tests for oxidizing adulterants on each primary specimen.
(d) You must perform additional validity tests on the primary specimen when the following conditions are observed:
(1) Abnormal physical characteristics;
(2) Reactions or responses characteristic of an adulterant obtained during initial or confirmatory drug tests (e.g., non-recovery of internal standards, unusual response); or
(3) Possible unidentified interfering substance or adulterant.
(e) If you determine that the specimen is invalid and HHS guidelines direct you to contact the MRO, you must contact the MRO and together decide if testing the primary specimen by another HHS certified laboratory would be useful in being able to report a positive or adulterated test result.
[69 FR 64867, Nov. 9, 2004]
READ MORESHOW LESS
['Drug and Alcohol Testing - DOT']
['Drug testing - Motor Carrier']
Load More
J. J. Keller is the trusted source for DOT / Transportation, OSHA / Workplace Safety, Human Resources, Construction Safety and Hazmat / Hazardous Materials regulation compliance products and services. J. J. Keller helps you increase safety awareness, reduce risk, follow best practices, improve safety training, and stay current with changing regulations.
Copyright 2024 J. J. Keller & Associate, Inc. For re-use options please contact copyright@jjkeller.com or call 800-558-5011.