As an employer, a BAT, or an STT, you must cancel an alcohol test if any of the following problems occur. These are “fatal flaws.” You must inform the DER that the test was cancelled and must be treated as if the test never occurred. These problems are:
(a) In the case of a screening test conducted on a saliva ASD or a breath tube ASD:
(1) The STT or BAT reads the result either sooner than or later than the time allotted by the manufacturer and this Part (see §40.245(a)(8) for the saliva ASD and §40.245(b)(8) for the breath tube ASD).
(2) The saliva ASD does not activate (see §40.245(a)(7); or
(3) The device is used for a test after the expiration date printed on the device or on its package (see §40.245(a)(1) for the saliva ASD and §40.245(b)(1) for the breath tube ASD).
(4) The breath tube ASD is tested with an analyzer which has not been pre-calibrated for that device's specific lot (see Sec. 40.245(b)(1) ).
(b) In the case of a screening or confirmation test conducted on an EBT, the sequential test number or alcohol concentration displayed on the EBT is not the same as the sequential test number or alcohol concentration on the printed result (see §40.253(c), (e) and (f) ).
(c) In the case of a confirmation test:
(1) The BAT conducts the confirmation test before the end of the minimum 15-minute waiting period (see §40.251(a)(1) );
(2) The BAT does not conduct an air blank before the confirmation test (see §40.253(a) );
(3) There is not a 0.00 result on the air blank conducted before the confirmation test (see §40.253(a)(1) and (2) );
(4) The EBT does not print the result (see §40.253(f) ); or
(5) The next external calibration check of the EBT produces a result that differs by more than the tolerance stated in the QAP from the known value of the test standard. In this case, every result of 0.02 or above obtained on the EBT since the last valid external calibration check is cancelled (see §40.233(a)(1) and (c)(3) ).
[67 FR 61522, Oct. 1, 2002; 71 FR 49384, Aug. 23, 2006]
READ MORESHOW LESS
['Drug and Alcohol Testing - DOT']
['Breath alcohol technician - Motor Carrier', 'Alcohol testing - Motor Carrier', 'Evidential breath testing device (EBT) - Motor Carrier']
Load More
J. J. Keller is the trusted source for DOT / Transportation, OSHA / Workplace Safety, Human Resources, Construction Safety and Hazmat / Hazardous Materials regulation compliance products and services. J. J. Keller helps you increase safety awareness, reduce risk, follow best practices, improve safety training, and stay current with changing regulations.
Copyright 2024 J. J. Keller & Associate, Inc. For re-use options please contact copyright@jjkeller.com or call 800-558-5011.