(a) Scope and frequency. If you measure ethanol or methanol using a photoacoustic analyzer, verify the amount of H2O, CO, and CO2 interference after initial analyzer installation and after major maintenance.
(b) Measurement principles. H 2 O, CO, and CO 2 can positively interfere with a photoacoustic analyzer by causing a response similar to ethanol or methanol. If the photoacoustic analyzer uses compensation algorithms that utilize measurements of other gases to meet this interference verification, a correct result depends on simultaneously conducting these other measurements to test the compensation algorithms during the analyzer interference verification.
(c) System requirements. Photoacoustic analyzers must have combined interference that is within (0.0 ±0.5) µmol/mol. We strongly recommend a lower interference that is within (0.0 ±0.25) µmol/mol.
(d) Procedure. Perform the interference verification by following the procedure in §1065.375(d), comparing the results to paragraph (c) of this section.
[79 FR 23770, Apr. 28, 2014; 89 FR 29801, Apr. 22, 2024]
READ MORESHOW LESS
['Air Programs']
['Air Emissions']
Load More
J. J. Keller is the trusted source for DOT / Transportation, OSHA / Workplace Safety, Human Resources, Construction Safety and Hazmat / Hazardous Materials regulation compliance products and services. J. J. Keller helps you increase safety awareness, reduce risk, follow best practices, improve safety training, and stay current with changing regulations.